Skip to product information
Neware CT-9000-5V15A Battery Testing System with High Speed Sampling (150 us), ENCT90005V15A

Neware CT-9000-5V15A Battery Testing System with High Speed Sampling (150 us), ENCT90005V15A

In Stock SKU: ENCT90005V15A
Use your own shipping account?

We support FedEx, UPS, and DHL third-party billing for institutional customers.

Place your order first, then email shipping@echemsupplies.com with your account details and order number. We'll generate the label using your account and refund your shipping charges, less a handling fee.

Review third-party shipping terms →


The Neware CT-9000-5V15A is the flagship high-performance cell tester in Neware's lineup. While the 4000 series is for general R&D and the 8000 series is for high-precision materials science, the 9000 series is designed for high-speed dynamic testing and ultra-fast response times. It is specifically engineered for researchers who need to simulate complex, real-world power demands—like the rapid current fluctuations in an electric vehicle or a power tool—on a single pouch cell.

CT-9000 can change its current output in less than 100 microseconds, it is the best tool for Dynamic Power Testing: (1) GSM/Telecom Pulses: It can simulate the rapid millisecond pulses required by mobile devices. (2) EV Drive Cycles (Standardized): You can import high-resolution "Current vs. Time" data from actual driving logs. The 9000 series can follow these jagged power curves with extreme fidelity, whereas slower testers would "smooth out" the peaks. (3) Fast DCIR: It can measure Internal Resistance using very short pulses (e.g., 10ms pulses), which helps separate pure Ohmic resistance from slower polarization effects.

Part Number
  • ENCT90005V15A
Power
  • AC220V±10%, single phase, 50/60Hz, 1500 W
Channel Numbers
  • 8 channels

Voltage Features


  • Charge/Discharge Voltage: 0 ~ 5 V
  • Minimal discharge voltage: 0 V (suitable for supercapacitor research)
  • Accuracy: ± 0.02% of F.S.
  • Stability: 0.01% of F.S.
Current Features
  • Current Range:

           Range 1: 0.015 mA - 15 mA

           Range 2: 15 mA - 150 mA

           Range 3: 150 mA - 1500 mA,

            Range 4: 1500 mA - 15 A

  • Accuracy: ± 0.02% of F.S.
  • Stability: 0.01% of F.S.
Output Power
  • 75 W/Channel
  • Stability: ± 0.01% of F.S.
Time
  • Current Response Time: ≤150 us
  • Working Step Time: ≤(365*24) h/step
    Time format-00: 00: 00.000 (h、m、s、ms)
Main Test Functions
  • Constant current charging/discharging (CCC/CCD)
  • Constant voltage charging/discharging (CVC/CVD)
  • Constant power charging/discharging (CPC/CPD)
  • Constant current constant voltage charging/discharging (CCCVC/CCCVD)
  • DCIR
  • Pulse Simulation Test
  • GITT

 

References:

S. Chen, et al., Critical Parameters for Evaluating Coin Cells and Pouch Cells of Rechargeable Li-Metal Batteries, Joule, 2019, 3, 1094-1105

Y. Son, et al., Analysis of Differences in Electrochemical Performance Between Coin and Pouch Cells for Lithium-Ion Battery Applications, Energy Environmental Materials, 2024, 7, e12615.

You may also like