Skip to product information
ECS-MB Medium Electrodialysis (ED) Testing System, CMBMEDTS

ECS-MB Medium Electrodialysis (ED) Testing System, CMBMEDTS

$8,999.00 USD
In Stock (Ships 3~8 Days) SKU: CMBMEDTS
Use your own shipping account?

We support FedEx, UPS, and DHL third-party billing for institutional customers.

Place your order first, then email shipping@echemsupplies.com with your account details and order number. We'll generate the label using your account and refund your shipping charges, less a handling fee.

Review third-party shipping terms →


A medium-scale (bench-to-pilot transition) Electrodialysis (ED) or Bipolar Membrane Electrodialysis (BMED) testing system sits squarely between a micro-volume desktop unit and a large industrial skid. It is engineered for advanced R&D, process optimization, and small-batch production validation—providing realistic hydrodynamic and mass-transfer data without requiring large fluid volumes.

The typical structural features of the medium electrodialysis testing system are: (1) Membrane Stack Capacity: Typically houses 10 to 30 cell pairs (compared to 5–10 in micro-bench units and 50–200+ in industrial stacks). Effective membrane active area per sheet ranges from 100 cm2 to 200 cm2 (e.g., standard 10 cm* 10 cm or 15 cm * 20 cm frames). (2) Processing Volume & Flow Rate: Designed for batch volumes of 10 L to 50 L per run, operating with recirculation flow rates between 1 L/min and 10 L/min to simulate real-world linear velocities and turbulence promoter efficiency. (3) Electrical Requirements: Typically paired with a DC power supply ranging from 0-30 V to 0-50 V at currents up to 10 A to 20 A, accommodating higher ionic loading than basic desktop cells.

The main application fields are: (1) Process Scale-Up & Kinetic Modeling: Generating reliable limiting current density curves, electrical resistance profiles, and specific energy consumption data (kWh/m3) needed to design full-scale industrial ED plants. (2) Organic Acid & Fine Chemical Deacidification: Purifying high-value fermentation broths, pharmaceutical intermediates, or food process streams by selectively removing inorganic salts without altering neutral organic components. (3) Targeted Ion Recovery & Wastewater Polishing: Concentrating specific heavy metals, mining leachates, or valuable alkali salts from industrial effluent streams down to strict discharge or recovery limits.

Part Number
  • CMBMEDTS (CMB-MEDTS)
Input Power/Voltage
  • AC220V±10%, single-phase, 50/60 Hz, 2000 W
  • Output DC voltage: 50 V; output DC current: 30 A

Testing Cells (Optional and Customizable, purchase separately)

  • (1) Testing cell with two chamber design (20cm*10cm) that can be used for the following conditions:

              (i) CEM + AEM mode

           

              (ii) BPM + CEM mode

             
               (ii) BPM + AEM mode
                    

           

  • (2) The testing cell with three chambers that is composed of bipolar membrane, AEM, CEM, polar membrane, and polar plate.

       

  • Chamber Material: Graphite manifold
  • Sealing Gasket: PTFE
  • Membrane Size: (1) 20cm*10cm (active area is 10cm*7.5 cm); (2) 30cm*14cm (active area is 15cm*10cm); (3) 40cm*20cm (active area is 20cm*15cm)
  • Larger sizes can be customized upon request
Water Tank Volume
  • 4 L
  • Minimal circulation volume is 1 L
Flow Pump
  • Flow pump speed: 1-50 mL/min
  • Four separated flow channes
  • Reservoirs and tubing are included for systematic testing
Control Unit 
  • PLC control system is used for integrating of operation functions
  • HMI touch screen for convenient operation 
Membranes for ED (Optional)
Dimension
  • L700mm * W600mm * H900mm
Weight
  • ~80 kg


References:

1. S. Abu Khalla, et al., Desalination via chemical energy: An electrodialysis cell driven by spontaneous electrode reactions, Desalination, 2019, 467, 418-427. 

2. M. Amores, et al. Electrodialysis as a Method for LiOH Production: Cell Configurations and Ion-Exchange Membranes, Adv. Sustain. Mater., 2025, 9, 2400402